				   MODEL 201
		    24-BIT DATA ACQUISITION SYSTEM















						    LAWSON LABS, INC.
						    3239 Phoenixville Pike
						    Malvern, PA  19355

							 610 725-8800
								or
							 800 321-5355


Rev. 7










			     TABLE OF CONTENTS


INTRODUCTION                                                  1

SECTION 1: INSTALLATION                                       2

SECTION 2: OPERATION                                          4
    Establishing Communications                               4
    Polled Mode Overview                                      4
    Software Calibration                                      5
    Connecting Analog Inputs                                  6
    Digital Input/Output Commands                             7
    Commands for Setting Up the A/D Converter                 7
    Other Commands                                            9
    Scanning Mode                                            10

SECTION 3: OPERATING SUGGESTIONS                             13
    Digital Input and Output                                 15


	    SECTION 4: USER'S TECHNICAL REFERENCE              16
    Polled Mode                                              16
    Output Commands                                          17
    Data Request Commands                                    18
    Special Commands                                         22

SECTION 5: TROUBLESHOOTING                                   25
			 
SECTION 6: INTERNAL ADJUSTMENTS                              26

SECTION 7: MODEL 201 SPECIFICATIONS                          26

APPENDIX                                                     29

WARRANTY                                                     30
				  INTRODUCTION


    The Lawson Labs, Inc. Model 201 24-Bit Data Acquisition
System is a high-resolution device for translating voltages
into digital form. The Model 201 communicates with a host
computer over an optically isolated RS-232 serial link. It has
6 multiplexed differential analog input channels with
provision for expansion to a total of 96 channels. There are
eight programmable gain ranges. Twenty digital input/output
lines are available for general purpose use. Future
expandability is provided through stacking option cards.
    The Model 201 is intended for DC and low frequency
applications. The data rate is programmable from under 1 to
over 300 samples/second. At a data rate of one sample/second,
the RMS noise approaches two counts, providing 23 bits
effective resolution. (The converter is guaranteed monotonic
to 24 bits.) Effective resolution decreases with increasing
data rates. Even so, 16-bit effective resolution is maintained
to over 250 samples/second.
    The Model 201 requires a single DC supply in the range of
11.5 to 50 volts. Current draw is typically 18 milliamps in
normal operation and only 2 milliamps in "sleep" mode. The
power and analog inputs are protected against substantial
overvoltages. These features, (plus error checking on the
serial link), make the Model 201 well suited for battery
powered remote operation.
    Both polled and scanning modes are available. In scanning
mode, the Model 201 maintains it's own time base and transmits
a pre-defined scan of up to 96 input channels at a preset
interval. Scanning rates from under 1 per day to over 300 per
second are supported. Other features under software control
include self-calibration, variable input filtration, and
bipolar or unipolar input ranges. 
    The exceptional resolution, stability, flexibility and
price are achieved by combining an accurate, but complex,
delta-sigma type A/D converter with a microcontroller
supervisor. The microcontroller makes it simple for the user
to enjoy the power of the delta-sigma converter by issuing
straight-forward commands over a standard serial interface.
								   

SECTION 1: INSTALLATION


    The Model 201 interconnections consist of three cable
connectors, a 32-pin expansion connector, and a 2-terminal
power connector. Locate the cable connectors at the bottom of
the card. They are for, from left to right, analog input,
serial input/output, and digital input/output.

NOTE:  Always handle circuit cards by the edges. Static
	 electricity can damage computer circuitry so care
	 should be taken to control static discharge.

    For operational checks, only the power supply and serial
cable need be connected. The power supply voltage can range
from 11.5 to 50 VDC and does not need to be regulated. Power
is connected to the terminals on the orange terminal block.
The power terminals are labeled "+" and "-". The wall-mounted
transformer supplied has a white stripe on the positive wire.
A battery, or other DC supply (in the 11.5 to 50 VDC range)
can be substituted. The board is protected against reverse
voltage but will not operate without a properly connected
supply. The Model 201 is well suited for battery power. In the
sleep mode only 2 mA is required.
    A supply voltage of greater than 24 VDC will increase
current consumption by as much as 3 milliamps. (The power can
be connected before or after the serial interface connection
is made.) 
    Figure 1 shows the input and output connectors. The 4-bit
optically isolated output allows use of Model 17, 20, and 35
multiplexers and multiplexed amplifiers. The pinout for the
expansion connector is given in Appendix 1.

    The serial interface uses a 9-pin connector and is
optically isolated from the host computer. The isolation
protects the host computer in the event of extreme
overvoltage. Note that the computer chassis ground is not
connected to ground at the Model 201. The serial pinout is
defined in Table 1.

    You will need to make the necessary connections for either
a 9- or 25-pin serial port (9-pin to 25-pin adapters may be
useful). Note that pins 2 and 3 are swapped in the cable for
25-pin serial ports. Some portable computers have 5 volt RS232
serial ports instead of 12 volt ports. If your computer has a
5 volt serial port you will want to move the jumper by the
serial connector (see Figure 2) to the 5 volt position.

    For maximum accuracy the board should be enclosed in a
shielded box. Open cell foam should be placed firmly against
both sides of the board to minimize air currents. Although a
copper/solder junction is not considered a good thermocouple,
there are many such junctions and they will have an effect at
the nanovolt level. (With a gain of 64, each count represents
9 nanovolts).
FIGURE 1: MODEL 201 BLOCK DIAGRAM 

	   Ŀ
	                   pinĿ 
	                    1  32-Pin Expansion Connector     
	                       
	    Ŀ                                    
	                                                      
11.5 toĴ Power Supply                                     
50 VDC                    Ŀ     Ŀ     
	                                    
	                         A/D                        
	    Ŀ                 Controller     
	    Input           Converter                     
	    Multiplexer                                   
	    with signal               
	    conditioning        Ŀ                      
	                        Jumper                      
	                                
	    Ŀ        Ŀ  Ŀ 
	    Analog             Serial   Digital In/Out    
	               
	   
                             
                                          Ŀ  Ŀ
                                                               
Six Differential    Input/Output   4-Bit Optically   8-Bit       8-Bit
Analog Input        to Host's      Isolated          Digital     Digital
Channels 0 thru 5   Serial Port    Output            Input       Output


		   TABLE 1: SERIAL CONNECTOR PINOUT

Description    Model 201      Computer Serial Port 
		   Pin Number     Pin Number
					9 Pin       25 Pin
----------------------------------------------------
   TX            2             2          3
   RX            3             3          2     
   DTR           4             4         20
   GND           5             5          7
   RTS           7             7          4

	   FIGURE 2: SERIAL PORT VOLTAGE JUMPER POSITION
			  
			   12                  12
			 Ŀ                    
			  x                  x  
			                   Ŀ                   
			  x                 x 
			                   
			   x                  x 
						   
			   5                   5

		  12 Volt Position     5 Volt Position
		  
SECTION 2: OPERATION

    The software driver provided with the unit includes an
executable file and a source file for QBASIC or Microsoft
QuickBASIC.  
    First make a backup copy of the support disk. You may want
to make a directory on your hard drive and copy the support
files into the directory. Your disk contains the following
files: 

IN_BASIC directory
DRIVR201.EXE  Executable driver software
DRIVR201.BAS  BASIC source code for above
DRIVR201.CFG  Configuration file
DEMO201.BAS   BASIC source code for polled mode demonstration
SCAN201.BAS   BASIC source code for scanning demonstration
SCRIPT.EXE    Executable file to generate sign-on scripts
README.SCR    Instructions for SCRIPT.EXE

WIN201 directory
    See the readme file in WIN201 for a description of the
Windows drivers and DLL.  

    Set the directory containing your support files as the
current directory. Type DRIVR201 from the DOS prompt. You
should see the communications menu on the screen. 

Establishing Communications

    Pressing the "1" key successively, cycles the baud rate
from 9600 through 300. The "2" key toggles the host COM port
between COM1 and COM2. By pressing the "3" key the user can
enter a buffer size between 1 and 32767 bytes. 512 bytes is
more than adequate for polled mode. A larger buffer may be
needed for scanning mode. Selecting option 4 begins a
reliability check for the serial communication link. The host
transmits characters which are echoed back by the Model 201.
The number of characters sent and the number of errors will
appear on the screen. If an error occurs a beep will sound.
When you are satisfied that communications are intact, press
"space" to continue. 
    If occasional serial errors are detected, select a slower
baud rate. If you get a "can't wake" message, double check the
power connections and serial cable pinout. 

Polled Mode Overview

    Once communications have been set up, the user can sign on
to the 201 by pressing the Enter key. After establishing
communications you will see the Polled Acquisition Screen.
Conversion results from the selected A/D channel will scroll
down the screen. The left hand number is the raw count from
the A/D converter. It will be a number between 0 and
16,777,215. The next number is the count translated into
volts. Next is the difference between the highest and lowest
reading. The next number is effective RMS resolution in bits.
The effective resolution is based on the last 20 conversions.
The last two numbers are the A/D channel selected and the
isolated external code selected. Commands specific to this
screen are listed at the bottom. Prompts, and responses to
commands are displayed to the right of the command line.
Bipolar or unipolar mode is indicated by a "+/-" or "+" with
the current gain setting in the upper right hand corner.
Pressing space resets the maximum and minimum readings.
Pressing the M key will display the command Menu, a summary of
commands. Enter brings back the polled acquisition Screen.
Commands can be issued from either the command menu or the
polled acquisition screen. If you are ever uncertain of the
current set-up, enter M to see the command menu screen.

Software Calibration

    At installation the standard values will have been
retrieved from the configuration file provided. To verify
function, press "C" and select A/D input Channel "7". Channel
7 is a dedicated zero channel. You should see a voltage in the
general vicinity of zero. Enter "O" to initiate an Offset
calibration. After the offset calibration the voltage should
read very close to zero. This offset calibration removes
errors caused by the input signal conditioning circuitry as
well as offset errors in the A/D converter itself. Confirm
that the reading is close to zero volts. Press "C" and select
A/D Channel "6". This channel is a dedicated full-scale
channel. Enter "F" to initiate a Full-scale calibration. After
the full-scale calibration the converter should read very
close to +5 volts. The full-scale calibration also removes
signal conditioning errors. 
    The "S" command (System calibration) performs the same
sequence automatically. It does the channel selection and both
calibrations and then reselects the original channel. Settling
times are accounted for. The "FILTER DELAY" message indicates
a settling time delay.
    Now you are ready to connect an input signal to one of the
A/D input channels.
    A battery is a convenient voltage source for checking the
Model 201. Connect the positive and negative ends of the
battery to a pair of positive and negative analog input pins
on the analog input connector. (Pins 1 and 9, respectively for
channel 0). You will also need a wire from one end (normally
the negative) of the battery to ground at the Model 201 to
insure that the input voltage at both input terminals is
within 6 volts of ground. The ground at pin 15 is provided for
the purpose. This extra ground is for common-mode requirements
only, and while necessary, it is non-critical. Remember that
the Model 201 is optically isolated and is floating compared
to the computer chassis ground. Enter "C" and select the input
channel to which you have connected the battery. Note that the
reading does not stabilize until the settling time has
elapsed. A typical D cell should read about 1.5 volts. Reverse
the wires and note the polarity change. Note that the plus
sign is implicit and does not appear. Connecting the input
wires directly together will cause a potential of zero volts.
An open circuit will read unpredictably. A positive
overvoltage will read 5 volts. A negative overvoltage will
read -5 volts. Severe negative overvoltages may appear to be
positive overvoltages. The analog input channels are protected
against continuous overvoltage up to 60 volts. With the
initial settings from the provided configuration file (10 Hz
data rate) the effective resolution should be 22 bits.
    Because of extreme resolution possible with the Model 201
it is necessary to carefully shield your input signals from
electrical noise. Electrical noise can be radiated through the
air and picked up by wiring and/or circuitry. It can also be
introduced via the power connections. Also, air currents can
create sufficient thermal effects to cause degradation of
effective resolution. 
    You can experiment now with any of the commands. You cannot
harm the Model 201 by issuing wrong commands. As long as you
do not answer Yes to the "DO YOU WANT TO SAVE YOUR
CONFIGURATION" prompt when you exit the program, you can
always restart from the last saved set-up. If no configuration
file is present, you will begin from the standard
configuration.
    It is possible to set up the Model 201 in a meaningless
way, i.e. if you do an offset and a full-scale calibration on
zero volts, then any voltage will read zero. If in doubt,
press "S" for system calibration. Also, with large filters
and/or averaging, you may have to wait for the response to a
command.

Connecting Analog Inputs

		FIGURE 3: ANALOG INPUT CONNECTOR PINOUT

	     PIN 8...+5V REF. OUTPUT
					     PIN 15...GROUND
	     PIN 7...REF. GROUND
					     PIN 14...CHANNEL 5-
	     PIN 6...CHANNEL 5+
					     PIN 13...CHANNEL 4-
	     PIN 5...CHANNEL 4+
					     PIN 12...CHANNEL 3-
	     PIN 4...CHANNEL 3+
					     PIN 11...CHANNEL 2-
	     PIN 3...CHANNEL 2+
					     PIN 10...CHANNEL 1-
	     PIN 2...CHANNEL 1+
					     PIN 9....CHANNEL 0-
	     PIN 1...CHANNEL 0+

Note:  For maximum protection, any unused inputs should be
	 connected to ground. This is done to protect the
	 circuitry from static discharges which can be of
	 extremely high voltage. Open inputs can also pick up
	 noise. Strain-relief is recommended for all permanent
	 wiring on the connector. Otherwise, physical stress may
	 cause the failure of an electrical connection. The
	 connector hood provided has a strain-relief clamp. 

Digital Input/Output Commands

    The "D" (Digital input) command reads an unlatched 8-bit
digital input word at pins 18 through 25 on the digital
input/output connector. The result is shown on the polled
acquisition Screen. The most recent result is also shown next
to the "D" command on the command screen. 

    The "E" command sends a 4-bit word to the optically
isolated External control code port at pins 1 through 4 of the
digital input/output connector. The current 4-bit word is
shown on the polled acquisition screen as the external control
code (EXT). These isolated outputs can only pull down in
relation to guard (pin 5). Pin assignments for the digital
inputs and outputs are shown in Figure 4. The four optically
isolated digital outputs can be used to control Lawson Labs
Model 17, 20, and 35 multiplexers and multiplexed amplifiers.

    The "U" (digital oUtput) command puts a latched 8-bit
digital word at pins 6 through 13 on the digital input/output
connector. After pressing "U", enter the word to output. 

	  FIGURE 4: DIGITAL INPUT/OUTPUT CONNECTOR PINOUT

    PIN 13...DIGITAL OUTPUT 7
						   PIN 25...DIGITAL INPUT 7
    PIN 12...DIGITAL OUTPUT 6
						   PIN 24...DIGITAL INPUT 6
    PIN 11...DIGITAL OUTPUT 5
						   PIN 23...DIGITAL INPUT 5
    PIN 10...DIGITAL OUTPUT 4                               
						   PIN 22...DIGITAL INPUT 4
    PIN 9...DIGITAL OUTPUT 3
						   PIN 21...DIGITAL INPUT 3
    PIN 8...DIGITAL OUTPUT 2
						   PIN 20...DIGITAL INPUT 2
    PIN 7...DIGITAL OUTPUT 1
						   PIN 19...DIGITAL INPUT 1
    PIN 6...DIGITAL OUTPUT 0
						   PIN 18...DIGITAL INPUT 0
    PIN 5...GUARD
						   PIN 17...GROUND
    PIN 4...OUTPUT D
						   PIN 16...GROUND
    PIN 3...OUTPUT C
						   PIN 15...GROUND
    PIN 2...OUTPUT B
						   PIN 14...GROUND
    PIN 1...OUTPUT A

Note:  Pins 1 through 4 are optically isolated digital
	 outputs. They function relative to pin 5 only.

Commands for Setting Up the A/D Converter 

    G) Gain - Gains of 1,2 ... 128 can be selected. Effective
resolution is reduced at gains above x4 or x8. The voltages
shown are after the gain stage. To get voltage at the input,
divide the displayed reading by the gain.

Note:  The 5 volt reference output cannot be used for full-
	 scale calibration at gains other than 1.


    W) Word length - The word length can be reduced from 24- to
16-bits using the W command. For higher data rates, maximum
throughput can only be obtained in the 16-bit mode. No
information is lost because the effective resolution is 16
bits or less at those high data rates. Note that not all of
the decimal places on the voltage display are significant in
16-bit mode.

    R) sample Rate - The A/D converter's data rate and
frequency response are set with this command. Rates from 10 to
1027 Hz are possible. There is a low-pass filter intrinsic to
the conversion process. The cut-off frequency of that filter
is the data rate times .262. For maximum effective resolution
use the lowest data rate that meets your needs.

    P) biPolar/uniPolar mode - the input range can be selected
as +/- 5 volts (bipolar) or 0-5 volts (uniPolar). Nominal
resolution improves in the unipolar mode for positive signals,
but some of that improvement is lost to fixed amplitude noise
sources.

    A) Average - The Model 201 will average consecutive
conversions. Press "A" and then "+" or "-" to cycle through
the choices. The number of conversions averaged must be a
power of 2, ie., 1, 2, 4, 8, 16, etc. The maximum number of
conversions to average is 32,768. Press Enter to register your
selection. The A/D data rate divided by the averaging factor
gives the number of samples transmitted per second. Note that
with high averaging values, data arrives slowly. At 10 Hz with
32768 averaging, each data point takes 54.6 minutes.

    T) filTering - The Model 201 has a programmable single-pole
filter in the signal path before the A/D converter. Press "T"
to cycle through the 3 choices for the filter cutoff
frequency: 400, 40, or 4 Hz. The lower filter cutoff
frequencies remove lower frequency noise, but require longer
settling times after channel changes.

    B) standBy - The standby command puts the A/D converter
into low-power mode. A/D conversions cannot be requested in
standby mode but all mode and calibration information is
retained. Issuing a "B" command a second time restores normal
operation.    

    S) System calibration - The "S" command does an automated
system calibration when the gain is set to one. A/D input
channel 7 is selected and an offset calibration is performed.
Then channel 6 is selected and a full-scale calibration is
done. Finally, the original channel is re-selected.
    For gains other than one, no system calibration will be
performed. A message indicating this fact will be displayed in
the lower right-hand corner of the screen. See below for a
description of how to calibrate your Model 201 for gains other
than one.
    Note that the Model 201 can remove offset or gain errors
from external circuitry up to 5% of full scale. Larger offsets
can be zeroed out but a reduced input range will result. Input
voltages over 5.1 volts will be clipped to protect the A/D.
    The most accurate calibration can be obtained using the
following method. Assume an amplifier is connected to A/D
input channel 0.

    1. Select A/D input channel 0.
    2. Apply zero volts to amplifier input. Assume the
	 amplifier output is +10 millivolts.
    3. Do an offset calibration. Channel 0 now reads 0 volts.
    4. Apply full-scale input signal to amplifier input.
	 Assume the amplifier output is now 4.9 volts.
    5. Do a full-scale calibration. A/D channel 0 now reads 5
	 volts. Calibration is complete.
    6. Note that now A/D input channel 7 will now read -10
	 millivolts and that channel 6 now will read +5 volts
	 overrange.

    Ground and 5 volts are brought out to the analog connector
to be used for cases where the system calibration command is
not sufficient. If the gain is greater than one, the 5 volt
reference signal cannot be used for full-scale calibration.
The best full-scale calibration results will be obtained if
the actual full-scale voltage is applied to an analog input
channel before issuing a full-scale calibration command for
that channel. The 5 volt reference and reference ground on the
analog input connector facilitate the derivation of a full-
scale calibration voltage for gains other than one. For
example, a 1.25 volt reference would be required for a gain of
four. Two series resistors with a ratio of 3 to 1 will provide
one-quarter of the 5 volt reference. The total impedance of
the divider network should be in the 5K to 200K range.


Other Commands

    Z) asleep (Z)/ awake (Z) - In sleep mode the A/D and input
section is shut completely off to minimize power consumption.
The control section checks periodically for serial activity
from the host. When activity is detected, the Model 201 turns
itself on and waits for start-up information. Note that for
lowest power consumption that the analog input voltages should
be disabled.

    K) checKsum - The Model 201 keeps a running total of all
serial data transmitted. The total is kept in modulo 256 form,
ie., any carry out of 8-bit binary is discarded. Therefore,
the checksum is always a number from 0 to 255. The host
computer can keep a similar total for all received words. The
host can request the Model 201's checksum any time
communications are open. The host can then compare the two
values to confirm that no serial errors have occurred. Both
checksums are set to zero after a checksum request. Checksum
OK is not a 100% guarantee that no errors have occurred
because multiple errors could cancel in the checksum. The more
frequently the checksum is requested, the smaller the chance
that a serial error will go undetected.

    C) Channel - Selects one of eight A/D input channels. The
inputs are differential and should always be maintained within
6 volts of the ground of the board for proper operation. Two
of the channels are dedicated to calibration.

Scanning Mode

    From the polled acquisition screen, a keypress of N will
take you to the scaNning menu. The scanning menu will allow
the operator to change the channel(s) to scan, scan interval,
and the type of scan. There are three types of scans
available, normal, self-calibration, and single-channel.

    Selections 0 through 5 correspond to the six A/D input
channels. With an external multiplexer connected, 1 to 96
channels can be selected for scanning. For normal and self-
calibrating scans, the A/D input channels are displayed with
the external control codes that are active for that channel.
In the single-channel mode only the selected channel will be
displayed with it's corresponding external control code. After
pressing the channel number the user will be prompted for the
external code for that channel. For multi-channel scans, both
the starting and ending external channels are entered. If the
ending channel code is below the starting code, that A/D
channel will be skipped during scanning. The single-channel
scan takes data from any one input channel. 

    Pressing 6 will prompt you for a new scan interval. The
minimum time is based on the baud rate, the number of readings
per scan, and the A/D speed. The A/D speed is determined by
the settings entered (press Esc then "M" to check your A/D
settings). The maximum time is determined solely by the baud
rate. The scan interval will always be an integer number of
clock counts, so the time resolution will increase with the
baud rate.

    The scan type is changed by the 7 key. In normal mode, all
of the selected A/D channels are scanned during each scan
interval. The calibration scan behaves in the same way with
the exception that a system calibration is done prior to each
scan. Because no settling time is needed, the highest sampling
rates are for single-channel scans. In normal and single-
channel scans, a system calibration is performed prior to the
first scan only.
Note: The self-calibration scan is only valid for a gain of
one.

    The enter key starts the scan and displays the status. The
screen will show the time the scan started, the interval time,
bytes per scan, elapsed time, scans expected, and scans
completed. Pressing the space bar will stop scanning at the
end of the current scan. Pressing Escape will cancel the
current scan. In either case, if any scans were completed,
they will be displayed next. Press space for subsequent scans.
Press Escape to return to the scanning menu.

    Escape from the scan menu takes you back to polled mode.

			 TABLE 2: COMMAND SUMMARY

  Function               Range/Description
------------------------------------------------------------------------
   "A" - Averaging       1 to 32768, incremented in powers of 2, ie., 1,
				   2, 4, . . .
   
   "B" - standBy         toggles between standby mode and normal mode
   
   "C" - Channel         analog input channels 0 to 5; channel 6 is +5 
				 volts reference for full scale calibration,
				 channel 7 is 0 volts for offset calibration
   
   "D" - Digital input   reads an unlatched 8-bit digital word at pins 18
				 through 25 on the digital input/output connector.
   
   "E" - External        sends a 4-bit word to the optically
     control code        isolated external control code port 
   
   "F" - Full-scale      makes the selected channel read 5 volts
   
   "G" - Gain            1, 2, 4, 8, 16, 32, 64, 128
   
   "K" - checKsum        requests a checksum from the Model 201   
   
   "O" - Offset          makes the selected channel read 0 volts
   calibration                     
   
   "P" - biPolar/        the input range can be selected as +/- 5 volts
   uniPolar mode           (bipolar) or 0-5 volts (unipolar).
   
   "R" - data Rate       10 Hz to 1 KHz
   
   "S" - System          calibration automatically performs an offset 
				 and full scale calibration
   
   "T" - filTer          400, 40, or 4 Hz cutoff frequency
   
   "U" - digital oUtput  puts a latched 8-bit digital word at pins 6
				 through 13 on the digital input/output connector. 
   
   "W" - Word length     16- or 24-bit conversion result
   
   "Z" - asleep (Z)/     In sleep mode the A/D and input section is shut
   awake (Z)             completely off to minimize power consumption. 
   
SECTION 3: OPERATING SUGGESTIONS

     The Model 201 A/D card features fully differential
inputs. A basic understanding of differential measurements
will help you to use your card to best advantage. The plus and
minus input pins should be wired directly to the voltage being
measured. In this way, it is assured that the only current
flowing in the wires will be the input current of the A/D
converter. Because the wires have finite resistance, any
current flowing will cause a voltage drop and a corresponding
error. The A/D card requires a vanishingly small input current
so the error caused by even very long wires is negligible. For
proper operation of the Analog Interface it is necessary that
the positive and negative inputs both be within 6 volts of
ground. For a floating voltage source, this is generally
accomplished by connecting a third wire between the Model 201
ground and a ground terminal at the source of the measured
voltage. Ground currents may flow in this wire, but the
resulting voltage drop will not cause a measurement error.
Redundant grounding can cause ground loops. Ground loops can
cause unpredictable behavior.

FIGURE 5: TYPICAL INPUT CONNECTIONS

		Ŀ                      Ŀ
    Ĵ  V  Ŀ                               
Ŀ               Ĵ+ 0                   
BATTERY      Ĵ-                     
       Floating voltage                             
      (Ground minus)                              
						                          
   Ŀ Ĵ+ 1                   
     V   Ĵ-                     
       Ground-referenced                  A/D         
	 (No second ground)               INPUTS        
	                                                       
    GROUND                     Ĵ+ 2                   
			    Ŀ   Ĵ-                     
	     Ĵ  V           .                   
                          .                   
           Ŀ              *                  
Ŀ  Ĵ  V  Ĵ+ 5                   
BATTERY                Ĵ-                     
                                   
                                     
                
     Multiple floating voltages        Power Supply Ground
	 (Ground single minus)              (or pin 15 on the
							analog input connector
			   * Remember to ground unused inputs

Best results are obtained with filtered, buffered voltages.
Electrical noise travels through the air and can be picked up
by interconnecting wires. The first defense against noise is
shielding. Use shielded wire with the shield connected at one
or both ends to ground. (See above). The lower the impedance
of the voltage source, the less susceptible the wiring will be
to electrical noise. If noise problems persist, try to locate
the source of the interference and shield it. Electric motors,
electric heaters and flickering fluorescent lights are
potential sources of interference.

 TABLE 3: EFFECT OF GAIN AND DATA RATE ON EFFECTIVE RESOLUTION

DATA
RATE:                 EFFECTIVE RESOLUTION, bits
--------------------------------------------------------------------
	    Gain   Gain   Gain   Gain   Gain   Gain   Gain   Gain 
	    of 1   of 2   of 4   of 8   of 16  of 32  of 64  of 128
10 Hz     22     22     22     21     20     19     18     17
25 Hz     20     20     20     20     19.5   18.5   17.5   16.5
50 Hz     19.5   19.5   19.5   19.5   19     18.5   17.5   16.5
100 Hz    19     19     19     18.5   18     17.5   17     16
250 Hz    15     15     15.5   15.5   15.5   15.5   15     14.5

* Effective resolution is defined as total resolution minus
RMS noise.

		  TABLE 4: APPROXIMATE SETTLING TIMES

				     SETTLING
				     TIME: sec
EFFECTIVE        Filter Cutoff    Filter Cutoff    Filter Cutoff
RESOLUTION        Setting of       Setting of       Setting of
(bits)             400 Hz            40 Hz            4 Hz
------------------------------------------------------------------
16                 .030             .30               3.0
20                 .035             .35               3.5
24                 .043             .43               4.3

Note: For maximum accuracy, wait twice the theoretic settling
times.

Digital Input and Output

    The digital input lines will accept signals from TTL or 5
volt CMOS logic. They will also work with contact closures to
ground. Inactive input pins are pulled up to 5 volts by 100K
resistors. The input voltage should not go above 5 volts or
below ground.

    The 8-bit digital output word is also 5 volt CMOS logic.
Each output can source or sink several milliamps.

    The optically isolated output port is primarily intended
for controlling external input multiplexers and multiplexed
amplifiers.

	 FIGURE 6: OPTICALLY ISOLATED DIGITAL OUTPUT CODE

  CODE           OUTPUT  D C B A

   0                     0 0 0 0
   1                     0 0 0 1
   2                     0 0 1 0
   3                     0 0 1 1
   4                     0 1 0 0
   5                     0 1 0 1
   6                     0 1 1 0
   7                     0 1 1 1
   8                     1 0 0 0
   9                     1 0 0 1
   10                    1 0 1 0
   11                    1 0 1 1
   12                    1 1 0 0
   13                    1 1 0 1
   14                    1 1 1 0
   15                    1 1 1 1

Electrically, each output line is the collector of an
optically coupled NPN transistor whose emitter is connected to
guard. Note that guard is not connected to ground on the Model
201. The guard is usually connected to the external device's
ground. Each output line can drive two low-power TTL loads or
many CMOS loads. An external pull-up resistor (from the
collector to the external positive supply) must be used.
Typically, 20 K ohms for low-power TTL or 47 K ohms for CMOS
is appropriate (the external supply can be as high as 24 VDC).
When the output code is 0, all four phototransistors are off
and no current is drawn through the transistors. If the input
to an inverting buffer is connected to output line A (with the
appropriate pull-up resistor), the input level is forced high
by the resistor so the buffer output would be 0. If the output
code is 1, the A transistor will turn on and pull the buffer
input to a logic level 0. The buffer output will then be a
logic level 1.

SECTION 4: USER'S TECHNICAL REFERENCE

    DRIVR201 software is intended to demonstrate the features of
the Model 201. It is not a general purpose data acquisition
package.
    For many applications, one configuration for the Model 201
will suffice. To simplify the task of including Model 201
support in existing software you will find a program called
DEMO201.BAS on your disk. This program will load a DRIVR201.CFG
configuration file and allow data collection in polled mode
only. You may find it desirable to use DEMO201 once you have
settled on a configuration for your Model 201. SCAN201.BAS
performs the same task as DEMO201.BAS except that it supports
scanning mode only.

    After power-up (or a reset) the Model 201 waits for sign-on
for approximately 8 seconds. If there is no serial input
activity, the Model 201 puts itself to sleep. To sign on, the
host sends a reset command at 300 baud. If the Model 201 is in
sleep mode or at a different baud rate it will not necessarily
respond correctly on the first attempt, so repeated tries may
be required. The response should be Hex 80 if the Model 201 was
asleep or Hex 3 if it was awake.
    After the host receives a Hex 3 it should pause and then
send the sign-on command (Hex 88) with a baud code. The Model
201 will respond by sending back the baud code at 300 baud. The
host and the Model 201 now switch to the chosen baud. 
    The integrity of the communication link can now be tested.
Any character sent to the Model 201 will be echoed back. The
host must wait for the echo before sending another character.
Echoing will continue unless the host sends nothing for
approximately 8 seconds. Then the Model 201 will transmit a
communications check error and put itself back to sleep.
    When the host is ready to proceed with sign-on it should
send a "null" (Hex 0). Now the Model 201 is ready to accept the
initialization values.

Note:  The host should set its running checksum for all
	 received data to zero at this point.

    Those initialization values are transmitted in four packets
of three characters each. The first two bytes in a packet are
data. The third is a checksum byte (the sum of the first 2
bytes modulo 256). The first two initialization packets contain
the following information: A/D data rate, gain, word length,
standby status, and bi/unipolar range. The next two packets set
averaging, pre-filtering, and polled/scanning status.
    The Model 201 will retransmit three mode words from the
first two packets for confirmation (these mode words are
actually written into the A/D converter and read back before
retransmission.

Polled Mode:

    You will normally sign on in polled mode. (See Page 20 for
scanning mode). After sign-on, the Model 201 will wait
indefinitely for instructions. 
    The instruction packets consist of three characters: a
command token, an argument, and a checksum. The argument
character can contain data from the host or it may be unused,
depending on the command. If the packet is not received
correctly, the Model 201 will transmit a single character error
code and wait for sign-on.
    There are three types of commands. Output commands send data
to the Model 201. Data request commands cause the Model 201 to
transmit characters back to the host.
    Output commands can be issued at any time after sign-on. The
host should wait for the response to a data request command
before sending another command.
    The third type of command is the single-byte command. There
are two single-byte commands: cancel and reset. Both supersede
any pending commands. That is, they take effect immediately.
The cancel command stops a data request command. The reset
command sends the Model 201 back to sign-on. If the host
doesn't sign on within approximately 8 seconds, the Model 201
goes to sleep. 
    Data requests with extensive averaging can take minutes to
complete. The cancel command will stop the pending data request
even if the Model 201 has begun to transmit the result.
Therefore, it is usually necessary to empty the serial input
buffer after issuing the cancel command.
    The reset command may also leave partially completed
transmissions in the buffer and should be handled similarly.

Output Commands:

    There are eight output commands. They can be sent at any
time after sign-on except when a data request command has been
issued but not answered.

FILTER COMMAND           Token = Hex 3

    Argument: 0 = 4 Hz cutoff
		  1 = 40 Hz cutoff
		  2 = 400 Hz cutoff

    Sets the single-pole pre-filtering cutoff frequency to the
    value specified by the argument.
	 
CONTROL CODE COMMAND     Token = Hex 1

    Argument: 7-bit channel selection code, 0-127

    The lowest 4 bits control optically coupled output bits A,
    B, C, & D on the 25-pin DB connector. The next 3 bits select
    the A/D analog input channel (0 through 7). The most
    significant bit is ignored.

AUXILIARY OUTPUT COMMAND        Token = Hex 2

    Argument: data 0-255

    The data word will appear at the digital input/output 25-pin
    connector. Data bit 0 is at pin 6, data bits 1-7 are at pins
    7-13. The data is latched and will remain until another
    command is issued (or power is cycled).

AVERAGE COMMAND          Token = Hex 4

    Argument: 0,1,2,3,4,...15

    The argument is the power of 2 of the number of points to
    average for each transmitted reading.

    Note that at the 10 Hz data rate if averaging is set to 15
    (215 = 32768) each reading will take over 54 minutes.

EXPANSION OUTPUTS        Tokens = Hex 6,7,8, & 9

    Argument: 8-bit data 0-255

    Writes data to an expansion card (if present). An expansion
    card is needed to take advantage of these commands.

Data Request Commands:

    After issuing these commands the host should wait for a
response before sending another data request. When a data
request command is sent in scanning mode, the data will be sent
immediately if a scan is not in progress. Otherwise, the data
will be sent at the end of the current scan. The only data
request commands that are valid during scanning are the read
digital input command and checksum command.
    The data request commands first retransmit the command
token, then send the requested data byte or bytes.

READ DIGITAL INPUT       Token = Hex 80

    Argument (Hex)
	OC      Expansion 0
	2C      Expansion 1
	4C      Digital Input
	8C      Expansion 3
	CC      Expansion 4

    The Model 201 will respond with the command token followed
    by one byte of digital input data. If the argument byte is
    4C, then the data will come from the 8-bit digital input
    port. An expansion card is needed to use the other digital
    inputs.

READ CONVERSION          Token = Hex 81

    Argument: Ignored

    Response: The command token is echoed immediately. Then the
    A/D is read and the 201 transmits two or three data bytes.
    The least significant byte is always sent first. To obtain
    the conversion result in counts:
    16-Bit mode   COUNT# = LOW + (MID * 256)
    24-Bit mode   COUNT# = LOW + (MID * 256) + (HIGH * 65536)

    To convert the count to millivolts use one of the following
    formulae.
    16-Bit unipolar mode MVOLTS =  COUNT# * .076294#
    16-Bit bipolar mode  MVOLTS = (COUNT# * .152588) - 5000
    24-Bit unipolar mode MVOLTS =  COUNT# * .0002980232#
    24-Bit bipolar mode  MVOLTS = (COUNT# * .0005960464) - 5000

    If the host changes the A/D mode or averaging during a read
    conversion, the result should be discarded.

OFFSET CALIBRATION       Token = Hex 82

    Argument: 7-bit channel selection code 0-127

    First, the Model 201 will echo the command token. Then, it
    will select the channel specified by the argument, wait the
    settling time and force that channel to read 0 volts. The
    Model 201 can zero offsets up to 1/2 of full scale. When the
    calibration is complete, a two- or three-byte calibration
    result is transmitted. The least significant byte is always
    sent first. Note that the offset channel remains selected
    after this command is issued.

FULL-SCALE CALIBRATION   Token = Hex 83

    Argument: 7-bit channel selection code 0-127

    First, the Model 201 will echo the command token. Then, it
    will select the channel specified by the argument, wait the
    settling time and force that channel to read 5 volts. The
    Model 201 can increase the gain by 15% or decrease the gain
    by 5% from the nominal value. Remember that the actual input
    voltage should not exceed +/- 5 volts. Also, if your offset
    calibration voltage was not zero volts, you will shift the
    available calibration range. The offset calibration should
    be done before the full-scale calibration. When the
    calibration is complete, a two- or three-byte calibration
    result is transmitted. The least significant byte is always
    sent first. Note that the full-scale channel remains
    selected after this command is issued.

    Note: If the gain is not 1, the desired full scale voltage
    should be applied to the input channel specified in the
    argument. The nominal full-scale voltages are x1 = 5; x2 =
    2.5; x4 = 1.25; x8 = .0625; x16 = .03125; x32 = .015625; x64
    = .0078125; x128 = .00390625.

    For gains above x1, if a full-scale reference is not
    available, use this sequence:

	 Set gain = 1
	 Do offset calibration (at channel 7)
	 Do full-scale calibration (at channel 6)
	 Reset desired gain
	 Do offset calibration (at desired channel)
	 
SET A/D MODE             Token = Hex 84

    Argument:     This command uses 3 argument bytes sent in
			sequence; MODEREGHI%; MODEREGMID%; and
			MODEREGLO%. A delay of several milliseconds is
			required before sending the arguments.

			Most of the A/D converter set-up information
			is compressed into these three bytes.
				     M
				     S
				     B
				   Ŀ
	  MODEREGHI%          M  M  M  G  G  G  X  S 
				    2  1  0  2  1  0       
				   
				   Ŀ
	  MODEREGMID%         W  X  X  P  X  F  F  F 
				    L             1 0 9  8 
				   
				   Ŀ
	  MODEREGLO%          F  F  F  F  F  F  F  F 
				    7  6  5  4  3  2  1  0 
				   
									   L
	  X   always 0                                     S
	  M   mode bits (normally zero)                    B
	  G   power of 2 for gain (zero for gain of 1)
	  S   1 = standby, 0 = normal operation
	  WL  1 = 24-bit word length, 0 = 16-bit word length
	  P   1= unipolar, 0 = bipolar
	  F   data rate = 19531.25/(F in decimal form)
		valid range for F in decimal is 19 to 2000

    See the DRIVR201.BAS source file for a complete description
    of the mode registers.

    First the command token is echoed. Then the three mode bytes
    are sent back to the host. The bytes are actually read back
    from the A/D converter before being re-transmitted.

VERSION                  Token = Hex 86

    Argument:     Ignored

    The Model 201 echos the command token followed by the
    version number. Additional features and custom commands may
    be available for particular version numbers.

CHECKSUM                 Token = Hex 87

    Argument:     Ignored

    The Model 201 sends back the command token followed by its
    running 8-bit checksum of all data transmitted as of when
    the checksum token was received. The host should zero it's
    checksum after confirming that the checksums match.
SCANNING MODE

    When the host signs on in scanning mode, the Model 201
requires 5 packets of additional information. The first 3 data
bytes in the packets are the scanning interval. The units for
that interval are dependent on baud rate. At 9600 baud each
scanning interval count is worth 256 microseconds. At 300 baud
each count is worth 8192 microseconds. The next 6 data words
are the starting and ending external control code for each A/D
channel. The most significant nibble is the starting code. The
least significant nibble is the ending code. If no external
multiplexer is present at that A/D channel, both the starting
and ending code should be zero. If a channel is to be skipped,
set the starting code greater than the ending code (normally
1,0). The seventh data word is unused. The Model 201 will take
care of all channel selection and filter delays for multi-
channel scanning.
    For single-channel scanning, the host must select the
desired channel before sending the start single-channel
scanning command.
    Any of the three start scanning commands can be sent now.
For each, the number of conversion data bytes will depend on
whether 16- or 24-bit mode is selected for the A/D. In either
case the low byte is sent first and the high byte last. The A/D
channels are scanned in ascending order. If external
multiplexing is used, all the external channels for an A/D
channel are transmitted before the next A/D channel is
selected.
    Output commands can be sent at any time during scanning.
Data request commands can be used sparingly. If a checksum or
read digital input is received by the Model 201 between scans,
it will respond immediately. If a scan is in progress, the
request will be handled after the completion of the scan. 

NORMAL SCAN              Token = Hex 89

    Argument:     Ignored

    When the Model 201 receives a normal scan token it echos the
    token back and sends a start of scan token (Hex F0) to the
    host. Next, the conversion results for all of the channels
    are sent followed by an end of scan token (Hex 0F). The 201
    will now wait the remainder of the scanning interval before
    sending another start of scan token. 

END SCAN                 Token = Hex 8A

    Argument:     Ignored

    This command will stop any type of scanning. If a scan is in
    progress, it will finish and then the Model 201 will echo
    back the end scan token.





SINGLE-CHANNEL SCAN      Token = Hex 8B

    Argument:     Ignored

    The Model 201 will echo the command token and then send data
    only. The host must pre-select the desired channel. There
    are no start or end scan tokens in single-channel mode.

SELF-CALIBRATE SCAN      Token = Hex 8C

    Argument:     Ignored

    Model 201 will echo the self-calibrate scan token. Then the
    Model 201 sends a start of scan token (Hex F0) followed by
    the results of a offset and full-scale calibration. The
    conversion results for all of the A/D channels are sent
    followed by an end of scan token (Hex 0F).
    

Special Commands:

SLEEP                    Token = Hex 88

    Argument:     Ignored

    The Model 201 echoes the sleep token then powers down the
    A/D converter and input conditioning circuitry. The digital
    outputs and optically isolated output port are not changed.
    To minimize power consumption in sleep mode, the host should
    zero these outputs before issuing sleep. A/D calibration and
    mode information is lost. 

    Note: The sleep token is also used in the sign-on sequence.

MASTER RESET             Token = Hex 0

    Argument:     NONE
    Checksum:     NONE

    The master reset is a single-byte command. Sending a master
    reset has the same effect as cycling power to the Model 201.
    The host must revert to 300 baud to re-establish
    communications.

CANCEL                   Token = Hex 85

    Argument:     NONE
    Checksum:     NONE                       
							   
    The cancel command is a single-byte command. It cancels any
    pending data request command. The Model 201 will echo the
    cancel token after receiving the cancel command.
	    
	    Ŀ
  > Open COM port with           FIGURE 7: SIGN-ON SEQUENCE
          DTR high and RTS low           (All data are in hexadecimal)
         Ĵ
           Wait .2 seconds      
           Send 0               
           Wait .2 seconds            ** Short sign-on for version 4 
                  and later
            ͻ
  <Ķ Is response 3?              See page 23 for variable
            ͼ            definitions
                      YES     
            Ŀ
             Wait .2 seconds 
             Clear Checksum  
             Send 88 (normal)
             or 99 (short)** 
            Ĵ
             Wait .1 seconds 
             Send BAUD% code 
            
           ͻ
  <Ķ Is Response BAUD%        Ŀ
           ͼ          Ŀ
                      YES                 Wait 2 seconds for Offset 
       Ŀ       and Full-Scale Calibration
        Open COM port at new baud       
        with DTR high and RTS low       Ŀ
              Empty Buffer              
            ͻ            Discard contents          
             Short Sign-On?     
            ͼ YES        Ŀ
                      NO                   Sign-On complete          
       Ŀ        Model 201 calibrated      
        Optional echo test                Polled mode               
        Send 0 to end echo test           rate = 10 Hz, gain = 1    
       Ĵ        averaging = 1, 24-bit data
        Send MODEREGHI%                   bipolar, filter = 40 Hz   
        Send MODEREGMID%                 
        Send Checksum             
       Ĵ   Ŀ
        Send MODEREGLO%                 Ŀ
        Send Placeholder                 Send scan interval low    
        Send Checksum                    Send scan interval mid    
       Ĵ       Send checksum             
        Send AVERAGE%                   Ĵ
        Send FILTER%                     Send scan interval high   
        Send Checksum                    Send CHAN0 external codes 
       Ĵ       Send checksum             
        Send Placeholder                Ĵ
        Send Mode (Polled or Scan)       Send CHAN1 external codes 
        Send Checksum                    Send CHAN2 external codes 
              Send checksum             
       ͻ      Ĵ
        Receive 3 bytes, should          Send CHAN3 external codes 
  Ķ match (MODEREGHI% and 1F)        Send CHAN4 external codes 
    NO   MODEREGMID%, MODEREGLO%          Send checksum             
	  ͼ      Ĵ
			     YES                 Send CHAN5 external codes 
		  ͻ             Send Placeholder          
		   Polled Mode?      Send checksum             
		  ͼ  NO         
		     YES  <
	    Ŀ
	     Sign-On complete      
	     Remember to calibrate 
	     before taking data    
	    
	    
FIGURE 8: POLLED MODE SEQUENCE
(All data are in hexadecimal)
	    Ŀ
	     Normal Polled Mode             Ŀ
	     Operation                       To calibrate              
	             
			    <Ŀ    Ŀ
	    Ŀ         Offset Calibration        
	     Send 1 command token           Send 82 command token     
	     Send control code              Send 70 for channel 7     
	     Send checksum                  (or alternate channel)    
	             Send checksum             
	    Ŀ        
	     Wait settling time            Ŀ
	             Wait for offset result    
	    Ŀ        
	     Send 81 command token         Ŀ
	     Send Placeholder               Full-Scale Calibration    
	     Send checksum                  Send 83 command token     
	             Send 60 for channel 6     
	    Ŀ         (or alternate channel)    
	     Receive three or four          Send checksum             
	     byte result                   
	            Ŀ
	    Ŀ         Wait for full-scale result
	     Scale in Volts                
	        
		   ͻ
		    Is gain 1?  Ŀ
		   ͼ  NO          Ŀ
			      YES                 Divide volts by gain      
			                         
			    <
	    Ŀ
	     Store and/or display  
	     result                
	    


SIGN-ON VARIABLES

     SIGNONTOKEN    88 hex    normal sign-on
			  99 hex    short sign-on (polled mode only)

     BAUD%          BAUD%          0    1    2    3    4    5
			  baud rate      9600 4800 2400 1200 600  300

     MODE REGISTERS   see SET A/D MODE command on page 19

     AVERAGE%       AVERAGE%       0    1    2    3  . . .  15
			  points to avg. 1    2    4    8  . . .  32768

     FILTER%        FILTER%                  0    1    2
			  cutoff frequency (Hz)    4    40   400

     MODE           0    scanning mode
			  1    polled mode

     SCAN INTERVAL  SCANINT = (3906.25/2^BAUD%)*(seconds-.99995)
			  where seconds is desired scan interval. SCANINT is a
			  triple precision unsigned integer with bytes of low,
			  mid and high.

     CHAN0 to CHAN5 External codes (for scanning only) 
			  High nibble is 4-bit external control code for first
			  point on CHAN
			  Low nibble is 4-bit external control code for last
			  point on CHAN
			  
SECTION 5: TROUBLESHOOTING


1)  Can't sign on.

    A. Try a slower baud rate.
    B. Make sure that power is connected properly to the 201.
    C. Check the communications port number and make sure you are
	 connected to the proper DB connector.
    D. Double-check your cable. If you are using a 9 to 25-pin
	 adapter, does it match the pinout described on page 3.
    E. Check that the serial port has DTR high and RTS low.

2)  Model 201 won't respond to a data request.

    A. Reduce averaging, you may not be waiting long enough.
    B. Remember to wait for the result from the previous request
	 before issuing another.

3)  Data is consistent, but wrong.

    A. Do a system calibration.
    B. Check the gain setting.
    C. Check uni/bipolar setting.
    D. Make sure another A/D input channel isn't badly over-
	 range.

4)  Data is noisy.

    A. Lower the data rate or increase averaging.
    B. Check shielding and grounding. Check that the DC common-
	 mode range of +/-6 volts is being observed.
    C. Make sure another A/D input channel isn't badly over-
	 range.

5)  Optically isolated outputs are not working.

	 Check connection to guard.

6)  Didn't receive expected data, and instead receive alternating
    5 hex and 80 hex from Model 201.

	 Model 201 has put itself to sleep. You must sign-on to re-
	 establish communications. Likely reasons for going to
	 sleep include power cycle, checksum error, or invalid
	 command.   
	 
SECTION 6: INTERNAL ADJUSTMENTS


    Hardware calibration is set at the factory and should never
need adjustment. The software should always be able to calibrate
to yield peak performance.

    There are two potentiometers on the board. The potentiometer
closest to the connectors is the common-mode rejection
adjustment. The other adjusts the reference voltage. Changing the
reference voltage has the effect of changing the gain.

    If you wish to reset the common-mode adjustment first,
connect the + and - input pins of a channel to a ground on the
analog input connector. Zero the channel by using the offset
command. Now remove the connection to ground and connect both
input pins to the 5 volt reference on pin 8 of the analog input
connector. Adjust the common-mode potentiometer for a reading of
zero. Repeat for best results. 

    The A/D gain is set by connecting a known voltage to an
analog channel. Do a system calibration then adjust the gain
potentiometer to obtain the desired reading. Repeat for best
results.

SECTION 7: MODEL 201 SPECIFICATIONS

TYPE:                      24-bit sigma delta (these converters have 
				     excellent noise rejection). Serially interfaced.

MONOTONICITY:              24 bits (up to 50 Hz data rate)

LINEARITY:                 0.002% of full scale typical up to 50 Hz data rate

DIFFERENTIAL INPUT RANGE:  +/- 5 volts or 0 to 5 volts

COMMON-MODE RANGE:         +/- 6 volts

DC COMMON-MODE REJECTION:  96 dB minimum

RESOLUTION:                16 or 24 bits, software selectable

INPUT IMPEDANCE:           100,000 megohms typical

PROGRAMMABLE GAIN:         1, 2, 4, 8, 16, 32, 64, or 128

ANALOG INPUTS:             6, fully differential, protected to 60 volts, 
				     expandable to 96 channels using optional 
				     multiplexers

DIGITAL INPUT:             8-bits, contact closure or 5-volt logic compatible

DIGITAL OUTPUT:            4 optically isolated lines plus 8 latched logic 
				     outputs

POWER REQUIREMENT:         +11.5 to +50 VDC regulated or unregulated

TYPICAL POWER CONSUMPTION: up to +24 VDC power supply voltage
					operation . . . . 18 milliamps
					standby . . . . . 14 milliamps
					sleep . . . . . .  2 milliamps
					add 3 milliamps to each at +50 VDC 
					
SIZE:                      5 x 6.5 x 1.2 inches

SPEED:                     The Model 201 has a programmable conversion rate 
				     which also determines the cut-off frequency of
				     the low-pass filter. Representative values are 
				     given below. Effective resolution is 24 bits 
				     minus RMS noise in bits. Effective resolution
				     may be reduced at gains above x4.
				     
DATA RATE:                        10 Hz      30 Hz      100 Hz     300 Hz
CUT-OFF FREQUENCY:                2.6 Hz     7.9 Hz     26 Hz       79 Hz
EFFECTIVE RESOLUTION (RMS BITS):  22          20         19          15
SETTLING TIME (SECONDS):          .4         .13         .04        .013

SERIAL INTERFACE:            RS232, optically isolated, full duplex. 
					 Programmable from 300 to 9600 Baud. Checksum 
					 transmitted on request.
					 
OTHER FEATURES:              Also under software control are signal averaging and variable input
					 filtration. The Model 201 can operate in 
					 polled or scanning mode.  In polled mode, 
					 individual readings are transmitted on request. 
					 In scanning mode, a pre-defined block of data 
					 is transmitted at a preset interval. Sleep mode 
					 drastically reduces power.
					 
OPTIONAL EXPANSION CAPABILITY: A stacking expansion board can hold
					   memory, additional A/D or D/A converters, 
					   or additional digital input and output. 
					    
CUSTOM MODIFICATIONS:         The Model 201 can be custom programmed to 
					  perform a wide variety of tasks. If you do 
					  not see a particular feature here, it may
					  still be available. Special versions of the 
					  microcode have been done to add a variety of 
					  features. Call for information. the Model 
					  201 can be optimized for process control, 
					  remote data-logging, chromatography, or 
					  other applications.

APPENDIX

EXPANSION CONNECTOR PINOUT

PIN 1      +12 VDC
PIN 2      -12 VDC
PIN 3      +7 VDC
PIN 4      -7 VDC
PIN 5      +5 VDC
PIN 6      -5 VDC
PIN 7      Analog Ground
PIN 8      RFS IN
PIN 9      RFS OUT
PIN 10     TFS IN
PIN 11     TFS OUT
PIN 12     +2.5 V REF
PIN 13     -2.5 V REF
PIN 14     +5 V REF  
PIN 15     A Filter
PIN 16     B Filter
PIN 17     Digital Ground
PIN 18     Digital  +5 VDC
PIN 19     Spare Select 4
PIN 20     Spare Select 6
PIN 21     Expansion Write
PIN 22     Expansion Read
PIN 23     D0 Expansion Bus
PIN 24     D1 Expansion Bus
PIN 25     D2 Expansion Bus
PIN 26     D3 Expansion Bus
PIN 27     D4 Expansion Bus
PIN 28     D5 Expansion Bus
PIN 29     D6 Expansion Bus
PIN 30     D7 Expansion Bus
PIN 31     E Analog In Channel Select
PIN 32     F Analog In Channel Select

LIMITED WARRANTY

     The Lawson Labs, Inc. Model 201 is guaranteed against
defects in materials and workmanship for a period of one year
from the date of delivery. Products must be returned to Lawson
Labs for warranty service. Contact Lawson Labs at 800 321-5355
for return authorization before returning anything for service.
     The above warranty is in lieu of all warranties express or
implied. Lawson Labs will not be liable for indirect or
consequential damages caused by any defect in this product. Some
states do not allow the limitation of consequential 
damages, so the above exclusion may not apply to you.